skip to Main Content
+86 15205060783 info@unuo-instruments.com
DIN Test Procedure-Complete Step-by-Step Guide (ISO 4649, DIN 53516)

DIN Abrasion Test Procedure: Complete Step-by-Step Guide (ISO 4649 & DIN 53516)

The DIN abrasion test procedure is the definitive international method for evaluating the wear resistance of vulcanized rubber, thermoplastic elastomers, footwear outsoles, conveyor belts, and industrial hoses. Governed by ISO 4649 (which replaced the original DIN 53516 standard), this test measures material loss by pressing a cylindrical specimen against an abrasive sheet mounted on a rotating drum over a fixed sliding distance of 40 meters (84 drum revolutions).

Unlike simple visual wear checks, the DIN method normalizes test results against a calibrated standard reference rubber to eliminate variations in sandpaper sharpness. Below is the complete, laboratory-ready operating procedure, covering specimen preparation, abrasive sheet calibration, testing execution, and volume loss calculations.

Part 1: Specimen Preparation and Conditioning

Accurate results begin with strict dimensional and environmental control before touching the machine.

  1. Cut or Mold Specimens: Prepare cylindrical test pieces with a diameter of (16 +/- 0.2) mm and a minimum thickness of 6 mm using a rotating hollow drill lubricated with water/soap solution, or by direct molding.

  • Thin Material Rule: If the sample (such as a thin shoe sole layer) is thinner than 6 mm, bond a sample piece (at least 2 mm thick) onto a base rubber cylinder with a hardness of at least 80 Shore A. Never abrade down to the adhesive bond line.

  1. Determine Required Sample Quantities:

  • Test Rubber: Prepare 4 specimens per material: 3 for mass loss runs and 1 intact specimen for density measurement (use 10 mass-loss specimens for referee or dispute testing).

  • Standard Reference Rubber (Compound No. 1): Prepare at least 2 to 3 specimens. One piece is used for abrasive sheet calibration, and a single standard reference piece can be reused for up to 3 consecutive test runs during a complete test cycle, provided it cools to room temperature between runs.

  1. Environmental Conditioning: Allow at least 16 hours between vulcanization or molding and testing. Condition all test specimens, standard reference rubbers, and the abrasive sheet at standard laboratory atmosphere ((23 +/- 2) C and (50 +/- 5)% relative humidity) in accordance with ISO 23529 prior to testing.

Part 2: Machine Setup and Abrasive Sheet Calibration

Before testing unknown materials, the abrasive cloth (typically P60 corundum grit) must be verified to ensure its cutting aggressiveness falls within the mandatory 180 mg to 220 mg mass loss window under Method A (non-rotating).

  1. Set to Non-Rotating Mode (Method A): Inspect the specimen holder (rub head). Ensure the gear mechanism is disengaged (remove the toothed gear wheel or install the smooth non-rotating ring) so the test piece slides across the drum without rotating on its own axis.

  2. Clean the Abrasive Drum: Brush the abrasive sheet thoroughly with a stiff bristle brush (or activate the integrated vacuum extraction) to remove any residual rubber dust from prior operations.

  3. Weigh the Standard Reference Rubber (m1): Weigh a conditioned Standard Reference Rubber No. 1 specimen on an analytical balance accurate to 0.001 g (1 mg). Record this initial mass as m1.

  4. Mount with Exact 2.0 mm Protrusion: Insert the standard rubber into the specimen holder chuck. Place the supplied metal thickness gauge plate (2.0 mm) against the chuck face so the rubber protrudes exactly (2.0 +/- 0.2) mm outside the holder. Tighten the locking nut securely.

  5. Apply the 10 N Vertical Load: Lower the swivel arm onto the drum and place the standard load weights onto the rub head to exert a vertical contact force of (10 +/- 0.2) N. (Note: Use 5 N only if specified for ultra-soft materials).

  6. Execute the Calibration Run (84 Revolutions):

  • Slide the specimen holder carriage to the starting position at the edge of the drum.

  • Set the digital revolution counter to 84 cycles (corresponding to an exact 40 m abrasion path on a 150 mm diameter drum).

  • Start the machine. Observe the holder for abnormal chattering or vibration; abort the run if vibration occurs. The machine stops automatically at 84 cycles.

  1. Re-weigh and Calculate Calibration Loss (m2): Remove the standard rubber, brush off loose crumbs, and weigh it immediately to 0.001 g (m2). Calculate the mass loss: delta m = m1 – m2

  2. Evaluate and Adjust Abrasive Sharpness:

  • Qualified (180 mg <= delta m <= 220 mg): The abrasive sheet is calibrated and ready for formal testing.

  • Too Dull (delta m < 180 mg): The sandpaper is worn out. Discard it, mount a fresh abrasive sheet onto the drum (ensuring a butt joint gap <= 2 mm), and repeat calibration from Step 2.

  • Too Sharp (delta m > 220 mg): The new abrasive cloth is overly aggressive. To blunt it safely: a) Release and remove the 10 N load weight from the abrasion head. b) Replace the rubber specimen with the steel/metal calibration block. c) Run the steel block across the drum for 1 to 2 passes (or run additional standard rubber passes if delta m was only slightly above 220 mg, e.g., 220-300 mg). d) Remove the metal block, re-install the 10 N load, clean the drum with a brush, and re-verify with standard reference rubber until mass loss stabilizes between 180 mg and 220 mg.

Part 3: Formal Testing Procedure (Interleaved Sequence)

Because abrasive paper wears slightly during testing, ISO 4649 requires pairing standard reference runs with test specimen runs. Every single run requires brushing the abrasive sheet clean beforehand.

Step 1: First Reference Run (delta mr1)

  1. Clean the calibrated abrasive sheet with a brush.

  2. Weigh the standard reference rubber (0.001 g accuracy), mount it with 2.0 mm protrusion, apply the 10 N load, and run 84 cycles.

  3. Re-weigh the standard rubber and record its first mass loss (delta mr1). Allow the reference rubber to cool to ambient temperature.

Step 2: First Test Sample Run (delta mt1)

  1. Brush the abrasive sheet clean.

  2. Weigh Test Specimen #1 (0.001 g accuracy) and record its initial mass.

  3. Mount Test Specimen #1 with 2.0 mm protrusion, apply the 10 N load, and run 84 cycles.

  • High-Wear Material Adjustment: If the sample loses more than 400 mg over 84 cycles, stop the machine at 42 cycles (20 m), reset the protrusion to 2.0 mm (ensuring remaining thickness is >= 5 mm), and complete the remaining 42 cycles. If mass loss exceeds 600 mg, terminate the test at 42 cycles (20 m) and multiply the recorded mass loss by 2.

  1. Remove Test Specimen #1, brush off debris, re-weigh (0.001 g), and record the mass loss (delta mt1). (Note: Never reuse a test specimen for a second run).

Step 3: Repeat Sequence Twice (Total 3 Complete Cycles)

  1. Brush the abrasive sheet clean after every run.

  2. Using the same piece of standard reference rubber (cooled to room temperature) and fresh test specimens (#2 and #3), repeat the alternating sequence twice more:

  • Run Standard Reference #2 -> Record delta mr2

  • Run Test Specimen #2 -> Record delta mt2

  • Run Standard Reference #3 -> Record delta mr3

  • Run Test Specimen #3 -> Record delta mt3

  • Smearing Check: If testing soft compounds that smear onto the sandpaper, check if the reference rubber mass loss drops by more than 10% after a test sample run. If it drops > 10%, Method A is invalid for that compound and Method B (rotating specimen holder) must be used.

Part 4: Density Measurement and Result Calculations

Because rubber compounds vary widely in filler content and specific gravity, comparing raw mass loss (mg) is misleading. Results must be converted into Relative Volume Loss (delta Vrel) or Abrasion Resistance Index (ARI).

  1. Measure Test Sample Density (rho t): Using the 4th untested specimen and a precision density balance (hydrostatic weighing method per ISO 2781), measure the specific gravity of the test compound in mg/mm3 (numerically equal to g/cm3) and record as rho t.

  2. Calculate Mean Mass Losses: Calculate the arithmetic mean for both the test rubber (delta mt) and the standard reference rubber (delta mr) across the three runs: delta mt = (delta mt1 + delta mt2 + delta mt3) / 3 delta mr = (delta mr1 + delta mr2 + delta mr3) / 3

  3. Formula A: Relative Volume Loss (delta Vrel) – Primary Standard Metric Relative volume loss represents the volume of test material worn away (in mm3) normalized to a standard reference mass loss of 200 mg:

delta Vrel = (delta mt x delta mconst) / (rho t x delta mr)

Where:

  • delta Vrel = Relative volume loss in cubic millimeters (mm3)

  • delta mt = Mean mass loss of the test rubber (mg)

  • delta mconst = Defined constant mass loss of the standard reference rubber (200 mg for Reference Compound No. 1 under Method A)

  • rho t = Density of the test rubber (mg/mm3)

  • delta mr = Mean mass loss of the standard reference rubber tested alongside the sample (mg)

(Interpretation: Lower delta Vrel = Better abrasion resistance. High-grade tire treads typically score 60-90 mm3, while standard footwear outsoles range from 100-180 mm3).

  1. Formula B: Abrasion Resistance Index (ARI) – Optional Percentage Metric When expressing durability as a percentage relative to the reference rubber:

ARI = [(delta mr x rho t) / (delta mt x rho r)] x 100%

Where rho r is the density of the standard reference rubber (mg/mm3). (Interpretation: Higher ARI% = Better abrasion resistance).

Quick Troubleshooting and Common Operator Mistakes

  1. Issue: Calibration loss > 220 mg

  • Root Cause: Fresh P60 abrasive sheet is overly sharp.

  • Corrective Action: Remove 10 N weight, blunt gently with the metal block, re-attach 10 N weight, and re-test.

  1. Issue: Calibration loss < 180 mg

  • Root Cause: Sandpaper grit is clogged or worn flat.

  • Corrective Action: Brush thoroughly; if still < 180 mg, replace the abrasive sheet immediately.

  1. Issue: Inconsistent or erratic mass loss

  • Root Cause: Forgot to clean drum between runs or wrong protrusion length.

  • Corrective Action: Brush drum after every single run and always verify 2.0 mm protrusion using the metal gauge plate.

  1. Issue: Specimen chatters or tilts

  • Root Cause: Locking nut is loose or sample height is < 5 mm.

  • Corrective Action: Tighten chuck nut firmly; ensure minimum height is >= 6 mm at start and >= 5 mm at end.

Back To Top